A low-temperature ultra-high-vacuum scanning probe microscope with in situ electronic transport capabilities: from macro to nano in 10 minutes
arXiv:2607.29099v1 Announce Type: cross Abstract: We have developed a low temperature (LT), ultra-high-vacuum (UHV) system that combines two complementary techniques, scanning probe microscopy (SPM) and electrical transport measurements, within a single platform. By providing simultaneous access to the atomic-scale surface landscape and the macroscopic device response of the same sample, the setup enables direct correlations between local structural/spectroscopic signatures and global electronic transport behavior in two-dimensional (2D) devices. The system allows experiments where atomic-scale modifications or controlled manipulations are performed while continuously monitoring their impact
…